Skip to main navigation Skip to search Skip to main content

The impact of statistical choice on variability and error in maximal respiratory mouth pressure testing

Research output: Contribution to journalArticlepeer-review

Abstract

Background: Various statistics have been used to assess repeatability and reproducibility of maximal inspiratory and expiratory mouth pressure measurements (PImax and PEmax, respectively). Few studies have examined the impact of rater experience on these values. The aims were to assess: 1) how choice of statistic impacted repeatability of PImax and PEmax; 2) the impact of rater experience on reproducibility of PImax and PEmax.

Methods: Thirty-one healthy adult males completed one familiarisation and two experimental visits. PImax and PEmax were assessed using a hand-held respiratory pressure meter by three raters with different testing experience. The intraclass correlation coefficient (ICC), coefficient of variation (CV), standard error of measurement (SEM), smallest detectable change (SDC) and limits of agreement (LoA) were determined across experimental visits for a given rater and between raters to determine repeatability and reproducibility of PImax and PEmax.

Results: For PImax and PEmax the: ICC was > 0.97 for reproducibility and 0.89-0.95 repeatability; CV was < 4 % for reproducibility and < 6 % for repeatability; SEM was 6-9 cmH2O for reproducibility and 8-15 cmH2O for repeatability; SDC was 16-26 cmH2O for reproducibility and 22-40 cmH2O for repeatability; LoA ratio bias was -2 to 5 % for reproducibility and -4 to 2 % for repeatability; LoA ratio error was 9-17 % for reproducibility and 19-25 % for repeatability.

Conclusion: Providing test administrators are trained, rater experience does not impact replicate values. However, choice of statistic impacts interpretation of PImax and PEmax variability and measurement error and reproducibility were better than repeatability.
Original languageEnglish
JournalExperimental Physiology
Early online date17 Jun 2026
DOIs
Publication statusEarly online - 17 Jun 2026

Keywords

  • measurement error
  • respiratory mouth pressure

Fingerprint

Dive into the research topics of 'The impact of statistical choice on variability and error in maximal respiratory mouth pressure testing'. Together they form a unique fingerprint.